Amprobe ACDC-100 Spécifications Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 36
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
22
Electrical Specications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.
(1) Voltage
ACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV, ACDC-54NAV
Function Range Accuracy*
DC V
99.99V
± (0.7% + 2dgt)
999.9V
AC V
99.99V
± (1.0% + 5dgt)
50 ~ 500Hz
999.9V
LPF
99.99V
50 ~ 60Hz ± (1% + 5dgt)
61 ~ 400Hz ± (5% + 5dgt)
999.9V
* DCV <1000dgt, add 6 dgt to the accuracy. ACV <1000dgt, add 3 dgt to the
accuracy.
Overload protection: 1000Vrms
Input Impedance: 3.5M // <100pF
AC Conversion Type: AC Conversions are ac-coupled, true RMS responding,
calibrated to the RMS value of a sine wave input. Accuracies are given for
sine wave at full scale and nonsine wave below half scale. For non-sine
wave (50/60Hz) add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
CF 3 @ 460V,460A (for ACD-53NAV, ACDC-54NAV),
280A (for ACD-50NAV, ACD-51NAV, ACDC-52NAV)
2 @ 690V, 690A (for ACD-53NAV, ACDC-54NAV),
420A (for ACD-50NAV, ACD-51NAV, ACDC-52NAV)
AC+DC Vrms Accuracy: same as ACV spec. +DCV spec.
(2) Current
ACD-50NAV, ACD-51NAV, ACD-53NAV
Function Range Accuracy
AC A
99.99A
50 ~ 60Hz ± (1.5% + 5dgt) **
61 ~ 400Hz ± (2% + 5dgt) **
599.9A / 999.9A*
LPF
AC A
0.10A ~ 99.99A
50 ~ 60Hz ± (1.5% + 5dgt) **
61 ~ 400Hz ± (5% + 5dgt) **
599.9A / 999.9A*
* ACD-50NAV, ACD-51NAV: 599.9A; ACD-53NAV : 999.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 34 35 36

Commentaires sur ces manuels